SCIFIT SXT7000 User Manual Page 43

  • Download
  • Add to my manuals
  • Print
  • Page
    / 65
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 42
43
5. Install new combo heart rate PCB in location shown in the drawing at the
top of this procedure. Note: The placement is very critical. The
antenna will not pick up a wireless transmitter if the placement is off.
6. Plug the hand grip cables (P2745 on SXT’s or P2775 on SX’s) into the
new combo heart rate PCB. Note: The top pin is not used. The cable
header with the “L” written on it plugs into the next two prongs of the
contact heart rate header. The next pin is not used. The cable
header with the “R” gets plugged into the bottom two prongs.
7. Plug the heart rate cable (P2309) back into the combo heart rate PCB.
Note: This plugs into the top three prongs of the heart rate to upper
PCB header. The bottom four (4) prongs are not used.
8. Plug the other end of the heart rate cable into the wireless heart rate
header on the upper PCB (P2160). See picture below.
9. Reconnect all cables to the upper PCB.
The ribbon cable plugs into the bottom ribbon cable header on the
right side of the upper PCB (as shown in the drawing).
The FITKEY ribbon cable (P2418) plugs into the FITKEY ribbon
cable header on the upper PCB. This header is located just above
the contact heart rate header (as shown in the drawing).
There are two (2) gray Telco cables. The cable from the 8VDC jack
affixed to the console back connects to the 8-pin, +8VDC power
header on the upper PCB. The other cable from the COMM jack
affixed to the console back connects to the 8-pin, C-Safe header.
10. Test operation using the contact grips then a chest strap. Note: Contact
Grips do not work in Heart Rate program.